Cold Chain Monitoring Template & Widgets

Configure, interpret, review, and sign off cold chain reports

Overview

The Cold Chain Monitoring report template provides a ready-made structure for reviewing temperature performance across a fleet of monitored devices. It combines fleet-level KPIs and a time-series chart with three cold-chain-specific widgets for compliance review, excursion assessment, and approval.

Component

Purpose

Fleet KPIs

Summarize the maximum, minimum, and average temperature across all included devices.

Cold Chain Monitoring Table

Provide a device-by-device, grouped or both compliance overview

Fleet time-series chart

Plot fleet minimum and maximum temperature against configurable lower and upper limits.

Cold Chain Excursion List

List individual temperature and/or humidity excursions for assessment and sign-off.

Sign-off

Provide a final approval area for a responsible reviewer.

Cold Chain Widgets

Cold Chain Monitoring Table

The Cold Chain Monitoring Table is the fleet overview. It evaluates every selected device and/or each hierarchy group, against configured temperature and/or humidity targets and presents the result using green, yellow, and red status colours.

Row and device identification options

  • Row layout. Display one row per device, one row per group, or groups with their devices underneath. Groups are taken from the device hierarchy configured in the report "Settings".

  • Device identification. Show the device name alone, or add the serial number and location description.

  • Additional columns. Enable or disable the available metric and device-health columns to match the intended audience. You can decide to display the columns:

    • In Range: Share of covered time spent inside the target range. Warning below 99% and critical below 95% by default; both thresholds are configurable.

    • Longest Excursion: The single longest uninterrupted period outside the target range, including its time window. Critical-from duration controls when an excursion makes the row non-compliant by itself. A shorter excursion is still displayed but does not affect the grade. A value of 0 leaves no tolerance.

    • Min/Max: Lowest and highest reading anywhere in the row. Values are coloured when they leave the target range. Uses the configured target minimum and maximum.

    • Coverage: How completely the metric was recorded during the report period. Warning below 98% and critical below 90% by default; both thresholds are configurable.

    • Battery: You can display device battery. Make sure the Semantic "Battery" is selected in the field.

Core measurement settings

Setting

Meaning

Sample validity

Define how long one sample remains valid. After this period, time is counted as missing rather than as a continuation of the last reading.

Target range

Define the minimum and maximum acceptable values. Values outside this range count toward excursions.

Humidity

Optionally add humidity with its own minimum and maximum target values when devices report humidity.

Why sample validity matters: Sample validity affects both data coverage and excursion calculation. If reporting stops during an excursion and the validity period expires, the gap is treated as missing data. One apparent long excursion is therefore reported as two shorter excursions.

A legend below the table summarises the target range, time-in-range thresholds, longest-excursion critical duration, coverage thresholds, sample validity, and status colors: green for OK, yellow for warning, and red for critical.

Look-around window (Time-range & Resolution of the report)

Cold Chain Excursion List

The Cold Chain Excursion List presents each qualifying excursion as a separate record. It is intended for detailed review, investigation, disposition, and sign-off by the responsible person.

Excursion calculation settings

Setting

Meaning

Semantic

Select temperature or humidity.

Sample validity

How long a sample remains valid. Missing data after the validity period splits an otherwise continuous excursion.

Target range

Minimum and maximum acceptable values; 2 °C to 8 °C are the default temperature limits in the template.

Minimum duration

Only excursions lasting at least this long are listed; the template default is 30 minutes.

Exposure

Enable Show exposure to add an exposure column. Exposure combines how far readings exceeded the configured limit with how long they remained outside the range, expressed in degrees Celsius per hour (°C·h).

Device information and review fields

Device identification can include the serial number and location description. The widget can also provide fields for documenting the outcome of the review:

  • Assessment outcome

  • Disposition

  • Record reference

  • Signer name

  • Sign date

  • Signature

Excursion IDs: important behaviour

  • The same real-world excursion receives a different ID in another report run, even when the report configuration and time frame are unchanged.

  • An excursion repeated in reports with overlapping periods receives a new ID in each run.

  • The ID remains the same only for the same report run, such as a back-office regeneration of that run with the same configuration.

  • If new historical data is backfilled, the enumeration can shift, which may change assigned IDs even during regeneration.

  • For durable audit references, store the excursion’s device, start and end timestamps, measured limits, and the generated report/run reference together—not the excursion ID alone.

Sign-off widget

The Sign-off widget provides a simple approval block intended for a printed PDF. A responsible person can confirm that the report, device status, and recorded excursions have been reviewed.

  • Title. The default “Approved by” label can be changed.

  • Name. Identifies the responsible reviewer.

  • Date and time. Records when the report was approved.

  • Signature. Provides space for the reviewer’s handwritten signature.

  1. Confirm the report period, included devices, device hierarchy, and output format.

  2. Verify the target ranges and sample-validity settings for every monitored semantic.

  3. Review fleet KPIs and the chart for broad temperature trends.

  4. Use the Cold Chain Monitoring Table to identify warning, critical, or unreliable rows.

  5. Open the Excursion List and assess duration, peak, exposure, and surrounding data for each event.

  6. Document the assessment outcome, disposition, record reference, and reviewer details.

  7. Complete the final Sign-off block after all exceptions have been reviewed.